The Industry’s Most Versatile Pump.
Trusted by Industrial Hygiene Professionals Worldwide.
Sensidyne / Products / Gilian Air Sampling / GilAir Plus Personal Air Sampling Pump (1 – 5,100 cc/min)
The Gilian® GilAir® Plus air sampling pump is designed for industrial hygiene professionals who require the ultimate in performance, versatility, and ease of use. It combines cutting-edge features with a user-friendly design to ensure accurate and reliable sampling data.
Reach out to Our Expert Team for Tailored Safety Solutions and Technical Guidance.
PERFORMANCE
Firmware
Current Version: 2.5.4 (6/20/17)
Flow Range
Constant flow 20 – 5,100 cc/min. without external adaptors
Constant pressure 1 – 5,100 cc/min without external adaptors
Flow Modes
Constant Flow or High and Low Flow Constant Pressure
Accuracy
Flow Display – 5% or 3 cc/min of set flow, whichever is larger
Constant Flow Control – 5% or 3 cc/min of set flow, whichever is larger
Constant Pressure Control – 10% of back-pressure
Back Pressure Capability
(8-hour runtime)
5000 cc/min. up to 12″ H2O back pressure
4000 cc/min. up to 20″ H2O back pressure
3000 cc/min. up to 30″ H2O back pressure
2000 cc/min. up to 30″ H2O back pressure
1000 cc/min. up to 35″ H2O back pressure
450 – 1000 cc/min. up to 40″ H2O back pressure
20 – 449 cc/min. up to 40″ H2O back pressure
Run Time
8 hour minimum
Flow Fault
If flow changes exceed 5% within back pressure specifications, fault notification appears. If fault exceeds 30 seconds, pump shuts down. Selectable automatic fault recovery allows the pump to attempt restart every 3 minutes for up to 10 times or to hold until manual intervention.
ENVIRONMENTAL
Temperature Ranges
Operating 32°F to 113°F (0°C to 45°C)
Storage -4°F to 113°F (-20°C to 45°C)
Charging 41°F to 104°F (5°C to 40°C)
Humidity Ranges
Operating 5-95 %RH, non-condensing
Storage 5-98 %RH, non-condensing
GENERAL
Display
Flow rate, sample time, and sample volume in actual conditions
Controls
Indicators Green and Red LED’s
Dimensions
4.3 x 2.4 x 2.4 inches (10.9 x 6.1 x 6.1 centimeters)
Weight 20.5 ounce (581 grams)
Warranty
2 years from original shipment, 5 years for pump keypad, and 1 year on rechargeable NiMH battery pack
ELECTRICAL
Power Options
Removable rechargeable nickel-metal hydride (NiMH) battery pack, optional alkaline battery pack, or DC power.
Battery Level Indicator
On-screen icon displays battery level.
Interface Connectors
All pump interface is carried out via the Dock or keypad.
Dock Functions
Battery charging, computer interface, and auto-calibration.
Charging Time
Under 3.5 hours.
APPROVALS
ISO 13137:2013 Compliant
Intrinsic Safety
US/Canada NRTL (FM):
Class I, Div 1 Groups A, B, C, D, E, F, and G; Class II, Div 1 Zone 0; Class III, T4
ATEX/IEC:
Ex – II 1 G; Ex ia IIC T4
CE
Applicable Directives:
ATEX 94/9/EC; EMC 2004/108/EC (RFI); LVD 2006/95/EC; RoHS Compliant
Gilian’s new PC application elevates everything about Gilian CONNECT to the next level, empowering you to manage your air sampling operations with ease and control.
Key Benefits:
• Save time and resources: Automate workflows, reduce manual data entry, and streamline compliance reporting.
• Gain deeper insights: Analyze data efficiently with advanced tools and visualize results with clarity.
• Make informed decisions: Confidently interpret data and make informed decisions based on accurate and complete information.
• Improved user experience: Enjoy a modern, intuitive interface designed for ease of use.
Key Features: